FILTRATION SYSTEM - 400 HEPA |
|
|
The 400 CFM HEPA Filtration system removes biological agents and harmful airborne particulates. The system delivers up to 400 CFM, working at max air flow provides min 12 air changes per hour for rooms ≤ 2000 cubic feet*, meeting flow requirements of the “CDC Guidelines for Environmental Infection Control in Healthcare Facilities”. *Note: Subject to Room Leakage.
|
| Includes: |
- Double element filter design to extend filter life
- Audible and visual indicators
|
|
- Wheels with handle for easy and safe movement
|
| |
|
| Double element filter design | |
Extends filter life while assuring effectiveness |
| Monitors positive or negative room/shelter pressure | |
Displays on digital readout |
| Monitors the pressure drop of the system filters | |
Indication of filter change |
| Audible and visual indicators | |
Alerts the user when the system is in an alarm or warning state |
| Vertical design | |
Minimizes footprint and optimizes floor space |
| Upper and Lower Alarm and Warning set points for each of the pressures monitored | |
| Data Port (Optional Software for data logging of room/shelter pressure and alarm/warning conditions) | |
| Can be used in conjunction with TVI’s infection control shelters or as a stand alone unit for rooms | |
| |
| Part Number | |
FS-ZB-04-1 |
| Capacity | |
0-400CFM |
| Casing Construction | |
Continuously welded, stainless steel enclosure |
| Preset Audible Warning Set-point | |
Preset at High is ± 03, Low is ± 01 inches of water after 10 second delay, latched until acknowledged (user adjustable) with software |
| Power Requirements | |
120 volt AC, 1 phase, 3.5 Amp |
| Dimensions | |
| |